• M-CDC
  • S-CDC


Performance of M-CDC at a glance:

Susceptibility to radiated interference

  • Compliant to IEC 1000-4-3 and EN 61000-4-3.
    Uniform Area: 1.5 x 1.5 m (L x H) at 3 m distance. Frequency range: 80 MHz to 3 GHz.

Radiated Interference

  • Pre-compliant from 30 MHz to 3 GHz at 3 m distance.

Conducted Measurements

  • Conducted Emission and Immunity measurements can be carried out.

Optionally an 18 GHz upgrade package is available

 

Purpose

The MEDIUM Compact Diagnostic Chamberä (M-CDC™) is developed with the focus on susceptibility testing.

The M-CDC combines conformity with an enhanced frequency range up to 3 GHz .

Consisting of modules, the installed chamber can be dismantled and re-erected elsewhere. Upgrading to a 3 m chamber is also possible.

Immunity

The M-CDC features superior performance for radiated immunity testing. The highest accuracy is achieved according to IEC 1000-4-3 and EN 61000-4-3 in a 1.5 x 1.5 m uniform plane at 3 m test distance.

The M-CDC is compliant with the above standards. In addition, it performs perfectly up to 3 GHz.

Emission

Radiated emission tests for development and quality control at a 3 m distance can be perfectly covered by the M-CDC. With its fully anechoic layout the time-critical height scan of the receive antenna is no longer needed.

Radiated Immunity

The accuracy achieved to IEC 1000-4-3 and EN 61000-4-3 in a 1.5 x 1.5 m at a 3 m test distance complies with these standards.

As required, 75 % of the 16 points are within the 0 ... + 6 dB accuracy range.


Fig. 1: Uniform plane 1.5 x 1.5 m to IEC 1000-4-3 and EN 61000-4-3

Typical calibration data of the uniformity over frequency range can be seen in Fig. 2.

Not included is the test system, e.g. amplifier, antenna, generator and computer.


Fig 2: Typical homogeneity of the M-CDC, 3 m measurement distance, 80 MHz - 1 GHz, uniform plane 1.5 x 1.5 m


Fig 3: Typical deviation from free space attenuation of the M-CDC™, 3 m measurement distance


Technical description:

  • Chamber dimensions: 7.3 x 3.1 x 3.0 m (l x w x h) approx.
  • Usable space inside: 6.7 x 2.8 x 2.7 m (l x w x h) approx. This is reduced to 6.4 x 2.2 x 2.4 m (l x w x h) approx. between the partially hybrid absorbers.
  • Fully galvanized modular shielding system of 2 mm steel.

Electromagnetic performance of shielding compliant with EN 50147-1 and MIL STD 285:


Frequency

Performance

Field component

10 kHz

60 dB

H

100 kHz

80 dB

H

1 MHz

100 dB

H

10 MHz

100 dB

E

100 MHz

100 dB

E

1 GHz

100 dB

PW

3 GHz

90 dB

PW


  • Modular ferrite tile lining on all six faces inside the enclosure completely including partially compact hybrid absorbers.
  • One personnel door of 900 x 2050 mm clear opening with mechanical actuation aid. The door is equipped with a switch to be connected to the interlock system for amplifier auto power off.
  • Six honeycomb inserts for ventilation.
  • Four illumination spots of 300 W.
  • One three-phase power supply filter, 4 x 16 A, for EUT and general power supply.
  • Electrical distribution with breakers, three single phase outlets, one three phase outlet, one switch for illumination.
  • One detachable access panels (AP) equipped with:

AP I

2 N connectors 50 W

2 BNC connectors 50 W

5 FSMA connectors

4 fastenings for data line filters

    • Fully walkable floor for EUT load up to 200 kg/m² covered with felt material.
    • Two cable outlets are integrated in the floor:

    Cable outlet I

    1 outlet 230 V, single-phase, 10 A

    Cable outlet II

    3 outlets 230 V, single-phase, 10 A

    1 outlet 230/400 V, three phase, 16 A

    • Integrated in the floor are two cable ducts which feed the necessary cables from the access panels to the cable outlets.
    • The facility is fully self-supporting.
    • The total weight is approx 7000 kg

The following Options are available

  • ·Upgrade package to enhance the performance up to 18 GHz. The package consists of high performance hybrid absorbers which are assembled on top of the existing ferrite tiles. The frequency range for radiated emission and immunity testing is improved up to 18 GHz. This kit reduces the clear internal space to 6.4 x 2.2 x 2.4 m (l x w x h) approx.

Frequency

Performance

Field component

18 GkHz

100 dB

PW

    • Turntable, free-standing, 80 cm high, max. load 100 kg, electrically driven including installation kit and IEEE controller.
    • Turntable with wooden plate, dia. 1.2 m, max. load 200 kg, floor-mounted version, electrically driven including installation kit and IEEE controllers.
    • Antenna tripod with remote controlled polarization tiltable from 0° to 90°. Height of antenna fixture: 1500 mm, overall height: 2400 mm including installation kit and IEEE controller interface.
    • Camera system for max. 200 V/m from 30 MHz to 18 GHz incl. installation kit and color monitor. Fiber- optic transmission.

Other options and variants are available upon request.

Maintenance

  • The Albatross Projects EMC test chambers are well developed products and need only very few maintenance. In normal use it is sufficient to clean the contact area of the door every three months. The contact springs of the door have to be replaced after approx. 30,000 operations.

Service

  • We have established a worldwide service network to provide the best possible after-sales service. For further information please contact Albatross Projects GmbH.

Quality

  • Certified Quality-Management system since 15.10.1993

ISO 9001

Reg No. 404.6/QM/10.93 (AA)


Performance of S-CDC at a glance:

Susceptibility to radiated interference

  • Compliant to IEC 1000-4-3 and EN 61000-4-3.
    Uniform plane: 1.0 x 1.0 m (L x H) at 1 m distance.
  • Frequency range: 80 MHz to 1 GHz.
    Compliant if EUT including cabling fits into this area.


Radiated Interference

  • Precompliant from 30 MHz to 1 GHz at 1 m or 2 m distance.

Conducted Measurements

  • Conducted emission and immunity measurements can be carried out.

Optionally an 18 GHz upgrade package is available

Purpose

The SUPER Compact Diagnostic Chamber (S-CDC) is developed with the focus on susceptibility testing.

The S-CDC combines excellent performance with low space requirements.

Consisting of modules, the installed chamber can be dismantled and re-erected elsewhere. Upgrading to a CDC or 3 m chamber is also possible.

Immunity

The S-CDC features superior performance for radiated immunity testing. The highest accuracy is achieved to IEC 1000-4-3 and EN 61000-4-3 in a 1.0 x 1.0 m at 1 m uniform plane test distance.

The S-CDC is compliant with the above standards as long as the equipment under test (EUT) including all cabling can be covered by this area.

Emission

Radiated emission measurements for development and quality control at 1 or 2 m distance can be perfectly covered by the S-CDC. With its fully anechoic layout the time-critical height scan of the receive antenna is no longer needed. An emission testing can be done within a very convenient timing.

Included as well in the S-CDC package are the correlation factors to convert the results achieved in an S-CDC to those of a 10 m site.

Radiated Immunity

Uniform plane of highest accuracy according IEC 1000-4-3 and EN 61000-4-3: 1.0 x 1.0 m at 1 m test distance. As required, 75 % of the 9 points are within the 0 ... + 6 dB accuracy.

The maximum possible test distance is 2 m.


Fig. 1: Uniform plane 1x 1m

Typical calibration data of the uniformity over frequency range can be seen in Fig. 2. It is validated for horizontal and vertical polarization at 1 m measurement distance.

Not included is the test equipment, e.g. amplifier, antenna, generator and computer.

Fig. 2: typical homogeneity of the S-CDC, 1 m measurement distance, 80 MHz - 1 GHz, uniform plane 1 x 1 m

Technical description:

  • Chamber dimensions: 4.6 x 2.5 x 2.75 m (l x w x h) approx.
  • Usable space inside: 4.33 x 2.23 x 2.33 m (l x w x h) approx.
  • Fully galvanized modular shielding system of 2 mm steel.

Electromagnetic performance of shielding compliant with EN 50147-1 and MIL STD 285:

Frequency

Performance

Field component

10 kHz

60 dB

H

100 kHz

80 dB

H

1 MHz

100 dB

H

10 MHz

100 dB

E

100 MHz

100 dB

E

1 GHz

100 dB

PW

18 GHz

100 dB

PW

    • Modular ferrite tile lining on all six faces inside the enclosure.

    • One personnel door of 900 x 2050 mm clear opening with mechanical actuation aid.
      The door is equipped with a switch to be connected to the interlock system for amplifier auto power off.
    • Two honeycomb inserts for ventilation.

    • Four illumination spots of 100 W.

    • One three-phase power supply filter, 4 x 32 A, for EUT and general power supply.

    • Electrical distribution with three single phase outlets, one switch for illumination and breakers.

    • Two detachable access panels (AP) equipped with:

AP I and APII each

2 N connectors 50 W

2 BNC connectors 50 W

7 FSMA connectors

4 fixation for data line filters

Fully walkable floor for EUT load up to 200 kg/m² covered with felt material.

  • Two cable outlets are integrated in the floor:
  • Cable outlet I

    1 outlet 230 V, single-phase, 16 A

    Cable outlet II

    3 outlets 230 V, single-phase, 16 A

    1 outlet 230/400 V, three phase, 16 A

    • Integrated in the floor are two cable ducts which feed the necessary cables from the access panels to the cable outlets.

    • The facility is fully self-supporting.

    • The total weight is approx. 4500 kg

The following options are available

  • Upgrade package to enhance the performance up to 18 GHz. The package is consisting of ultra compact high performance hybrid absorbers which are assembled on top of the existing ferrite tiles. The frequency range for radiated emission and immunity testing is improved up to 18 GHz. The usable space inside reduces to 4.10 x 2.00 x 2.20 m (l x w x h) approx.

  • Turntable, free-standing, 80 cm high, electrically driven including installation kit and IEEE controller.· Antenna tripod with remotely controlled polarization tiltable from 0° to 90°. Height of antenna fixture: 1150 mm, overall height: 2000 mm including installation kit and IEEE controller interface.

  • Camera system for max. 200 V/m from 30 MHz to 18 GHz incl. installation kit and color monitor. Fiber- optic transmission.


Maintenance

  • The Albatross Projects EMC test chambers are well developed products and need only very little maintenance. During normal use an annual check of the door is sufficient. The contact springs of the door have to be replaced after approx. 30,000 operations.

Service

  • We have established a world wide service network to provide the best possible after-sales service. For further information please contact Albatross Projects.

Quality

  • Certified Quality-Management system since 15.10.1993

ISO 9001

Reg No. 404.6/QM/10.93 (AA)